Yokogawa YTA110 YTA 310 YTA320 Temperature Transmitter 4 to 20mA DC with digital communication HART protocol G1/2 1/2NPT M20
Features: Temperature Transmitters YTA110 YTA310/YTA320
Outstanding performance
Microprocessor-based sensing technology ensures high accuracy and stability.
High reliability
Dual-compartment housing realizes high resistance capability to harsh environments.
SIL Certification
YTA series temperature transmitters have been certified by TÜV NORD as being compliant with the requirements of the IEC61508 for SIL 2.
Variety of sensor inputs
The type of sensor input is user-selectable: thermocouples, RTDs, ohms, or DC milivolt.
Digital communication
The BRAIN and HART communication protocols are available.
The instrument configuration can be changed by the user with a BT200 terminal or HART communicator.
The YTA320 temperature transmitter also supports FOUNDATION fieldbus communication protocol.
Self-diagnostics function
Continuous self-diagnostics capability ensures long-term performance and lower cost of ownership.
LCD display with bargraph
The LCD display simultaneously displays a digital readout and a bargraph.
Dual universal inputs (Model YTA320)
The YTA320 can accept two thermocouple, RTD, ohm, or DC milivolt inputs. Differential or average temperature transmitter measurement can be selected. A sensor backup function automatically switches-over from the primary sensor to a backup if the sensor fails.
Accuracy (typical) | YTA110: ±0.1% of span YTA310/320: ±0.05% of span |
Cold junction Compensation accuracy | (For T/C only) ±0.5 deg C (0.9 deg F) |
Power supply effect | ±0.005% of calibration span per volt |
Ambient temperature effect | ±0.1%/10 deg C (YTA110) |
Input signal | YTA110/YTA310: single input, YTA320: dual input Input type is selectable: thermocouples, 2-, 3-, and 4-wire RTDs, ohm and DC milivolt |
Output | Two wire 4 to 20 mA DC or FOUNDATION fieldbus signal (YTA320) |
Supply voltage | BRAIN and HART: 10.5 to 42 V DC (10.5 to 30 V DC for Intrinsically safe type) Fieldbus: 9 to 32 V DC (9 to 24 V DC for Entity and 9 to 17.5 V DC for FISCO) |
Ambient temperature | Option code may affect limits -40 to 85 deg C (-40 to 185 deg F) -30 to 80 deg C (-22 to 176 deg F) with integral indicator |
Mounting | Optional mounting bracket can be used either for two-inch pipe or flat panel mounting. |
Weight | 1.2 kg (2.6lb) without integral indicator and mounting bracket. Integral indicator weighs 0.2kg (0.4lb). |
Degrees of protection | IP66, IP67 and NEMA 4X |
Certificates | CENELEC, CSA, FM, IECEx or TIIS |
SIL Certificate | Certified by TÜV NORD in compliance with the IEC61508 SIL 2 capability for single transmitter use. |